Las Piñas General Hospital and Satellite Trauma Center invites bids for two electrocautery units under a PHP 3,400,000 approved contract budget. The public notice specifies monopolar and bipolar modes, output ranges, touchscreen controls, a heavy-duty trolley, compatibility with existing Karl Storz laparoscopic towers and hospital consumables, a three-year warranty, five-year parts availability after warranty, a local repair facility and a loaner during repairs. Delivery is listed as 60 days. Offers close on 5 November 2026 at 9:00 AM as printed; no timezone is stated. A virtual pre-bid conference is listed for 22 October at 9:30 AM.
Supply of two electrocautery machines for the hospital.
The offered unit must support the specified monopolar/bipolar functions, output ranges and accessories.
Three-year warranty; parts and consumables available from an authorized Philippine service centre for five years after warranty.
Local repair facility and technician/biomedical engineer response within 48 hours, excluding weekends and holidays, during warranty; provide a loaner during repairs or maintenance.
Demonstrate compatibility with the hospital’s existing Karl Storz laparoscopic towers and consumables; equivalent non-proprietary supplies must not compromise function, safety or warranty.
Lots
Electrocautery machine
Tender details
Buyer
Las Piñas General Hospital and Satellite Trauma Center
Country
Philippines
Procedure
Not stated in the notice
Classification
Hospital medical equipment
Notice reference
13281606
Approved contract budget
PHP 3,400,000
Quantity
2 units
Delivery period
60 days
Submission deadline
5 November 2026, 9:00 AM as printed; timezone not stated
Pre-bid conference
22 October 2026, 9:30 AM, virtual via Zoom
What to check in the tender documents
The invitation does not state a timezone for the 9:00 AM closing time.
ELECTROCAUTERY MACHINE
ELECTROCAUTERY MACHINE
Specification:
- Heavy duty with monopolar and bipolar modes
- With at least 5 inches LCD touch screen control and one key recover button or its equivalent
- Configuration can be preset and saved according to the users’ habit
- Support the connection of adult per neonatal type neutral electrodes and split/non-split neutral electrode
- Power Supply: 110V/240VB AC, 50-60Hz (standard electrical configuration)
- Control modes: Manual and foot operated and wireless foot operated options
- High frequency: at least 300 kHz
- Power output adjustable power levels: 0 to 120W for monopolar, at least 70W for bipolar mode, at least 400W for monopolar-cut, at least 150W for monopolar-coag, at least 100W for bipolar-cut, 100 watts for bipolar-coag, pure-cut 0-300W, blend-cut 0-200W, and coagulation 0-120W
- With cutting mode, blend mode, spray mode and coagulation mode
- Dimension: Length=at least 38cm; Width=at least 31cm; Height=at least 12cm; Weight=at least 5kg
- With heavy duty trolley equipped with a heavy duty casters and secure wheel locks for safe, stable and easy mobility
Standard requirements:
- Bidder’s certificate that parts and consumables shall be available at the authorized Philippine service center/s for five (5) years after the warranty period
- Three (3) years warranty on parts and services and provision of appropriate calibration tools for preventive maintenance
- Must provide operating manual
- The proponent must have a local repair facility and availability of a service technician and/or biomedical engineer within 48 hours (except weekends and holidays) upon receipt of notice in case of equipment breakdown during the warranty period
- With a loaner unit to ensure uninterrupted operations whenever the equipment is under repair or maintenance
- Bidder must have at least 5 service Engineer/Technical with a National Certificate II in Biomedical Equipment servicing
- Compatibility with existing Laparoscopic Towers
- The electrosurgical unit must be fully compatible with the existing laparoscopic tower systems currently installed at the Las Piñas General Hospital and Satellite Trauma Center Operating Room Complex
- The supplier shall provide documentary evidence or certification of compatibility with the following existing laparoscopic tower brand (Karl Storz)
- Compatibility with existing consumables
- The electrosurgical unit must be compatible with the hospital’s existing electrosurgical consumables to avoid additional procurement costs and ensure continuity of service
- Compatibility shall include, but not limited to:
- Existing disposable and reusable electrosurgical pencils (electrocautery pens)
- Existing patient return electrodes (electrocautery pads), both split and non-split types currently in use
- The supplier shall provide written certification and demonstrate compatibility during product evaluation and acceptance testing
- No propriety lock-in
- The offered unit shall not require exclusive proprietary consumables if equivalent existing hospital consumables are available and compatible
- Use of existing hospital stocks of electrocautery pads and electrocautery pens shall not compromise the functionality, safety features, or warranty of the equipment
Line Items
Item No.
Product/Service Name
Description
Quantity
UOM
Budget (PHP)
1
Electrocautery Machine
Rebid
2
Unit
3,400,000.00
Pre-bid Conference
Date
Time
Venue
22/10/2026
9:30 AM
VIRTUAL PREBID CONFERENCE THRU ZOOM LINK:
https://us06web.zoom.us/launch/jc/83005793042
Meeting ID: 830 0579 3042
Passcode: PREBID1234