title: "Scanning electron microscope for quantitative analysis of geological samples"
canonical: "https://zephior.com/tenders/342e052b-a33d-4796-9918-94ca9c7f43f6/scanning-electron-microscope-geological-samples-eth-zurich"
last-updated: 2026-10-11T23:26:42.614017+00:00
---

# Scanning electron microscope for quantitative analysis of geological samples

ETH Zürich is seeking a replacement analytical scanning electron microscope for its Institute of Geochemistry and Petrology. The instrument must support quantitative EDS with reference materials and full matrix correction, include a field emission gun and SE/BSE detectors, and accommodate at least three petrographic thin sections and six one-inch epoxy mounts. Automated quantitative runs longer than 24 hours are required; a cathodoluminescence detector is optional. Delivery has strict lift, floor-loading and crate-size limits. This open, single-lot supply tender accepts electronic bids through Simap. The deadline was extended to 11 November 2026 at 12:00 CET. Detailed specifications and evaluation criteria are in documents that were not accessed.

- Status: open
- Country: CH
- Buyer: ETH Zürich - Abt. Procurement & Export Services
- Submission deadline: 2026-11-11T12:00:00+01:00 (instant)
- Original notice: https://ted.europa.eu/en/notice/-/detail/676851-2026
- Source checked: 2026-10-11T23:24:32.306399Z

## Source-backed public facts

```json
{
  "id": "342e052b-a33d-4796-9918-94ca9c7f43f6",
  "revision": 1,
  "modifiedAt": "2026-10-11T23:26:42.614017+00:00",
  "sourceCheckedAt": "2026-10-11T23:24:32.306399Z",
  "stale": false,
  "url": "https://zephior.com/tenders/342e052b-a33d-4796-9918-94ca9c7f43f6/scanning-electron-microscope-geological-samples-eth-zurich",
  "locale": "en",
  "title": "Scanning electron microscope for quantitative analysis of geological samples",
  "summary": "ETH Zürich is seeking a replacement analytical scanning electron microscope for its Institute of Geochemistry and Petrology. The instrument must support quantitative EDS with reference materials and full matrix correction, include a field emission gun and SE/BSE detectors, and accommodate at least three petrographic thin sections and six one-inch epoxy mounts. Automated quantitative runs longer than 24 hours are required; a cathodoluminescence detector is optional. Delivery has strict lift, floor-loading and crate-size limits. This open, single-lot supply tender accepts electronic bids through Simap. The deadline was extended to 11 November 2026 at 12:00 CET. Detailed specifications and evaluation criteria are in documents that were not accessed.",
  "status": "open",
  "recordedStatus": "open",
  "sourceStatus": "open",
  "country": "CH",
  "buyer": {
    "id": "eth-zurich-abt-procurement-export-services",
    "name": "ETH Zürich - Abt. Procurement & Export Services"
  },
  "category": {
    "de": "Rasterelektronenmikroskope",
    "en": "Scanning electron microscopes",
    "fr": "Microscopes électroniques à balayage",
    "id": "scanning-electron-microscopes"
  },
  "deadline": {
    "at": "2026-11-11T12:00:00+01:00",
    "date": null,
    "source_ref": "official_ted_xml#/ContractNotice/ProcurementProjectLot/TenderingProcess/TenderSubmissionDeadlinePeriod",
    "sourceRef": "official_ted_xml#/ContractNotice/ProcurementProjectLot/TenderingProcess/TenderSubmissionDeadlinePeriod",
    "collectorValue": "2026-11-10T23:00:00+00:00",
    "precision": "instant",
    "displayTimeZone": "visitor browser time zone for exact timestamps; date-only deadlines retain their calendar date",
    "semantics": "at or date is the reviewed deadline from the cited source. collectorValue is what the collector parsed; it can show a date-only deadline as local midnight, so prefer at or date when they differ."
  },
  "originalSource": {
    "url": "https://ted.europa.eu/en/notice/-/detail/676851-2026",
    "title": "Rasterelektronenmikroskop zur Bildgebung und quantitativen Analyse an geologischen Proben",
    "issuer": "ETH Zürich - Abt. Procurement & Export Services",
    "status": "open",
    "language": "en",
    "location": "CHE",
    "deadline_at": "2026-11-10T23:00:00+00:00",
    "description": null,
    "country_code": "CH",
    "published_at": "2026-09-30T22:00:00+00:00",
    "contract_type": null,
    "deadline_date": null,
    "procedure_type": "open",
    "reference_numbers": [
      "676851-2026"
    ],
    "classification_codes": [
      {
        "code": "38511000",
        "system": "CPV",
        "provenance": "source_declared"
      },
      {
        "code": "38000000",
        "system": "CPV",
        "provenance": "source_declared"
      }
    ]
  },
  "sourceName": "TED",
  "sources": [
    {
      "ref": "canonical_snapshot",
      "url": "https://ted.europa.eu/en/notice/-/detail/676851-2026",
      "kind": "canonical_snapshot",
      "retrieved_at": "2026-10-11T23:23:45.986621Z"
    },
    {
      "ref": "official_ted_api",
      "url": "https://api.ted.europa.eu/v3/notices/search",
      "kind": "official_notice_api",
      "retrieved_at": "2026-10-11T23:24:31.898307Z"
    },
    {
      "ref": "official_ted_xml",
      "url": "https://ted.europa.eu/en/notice/676851-2026/xml",
      "kind": "official_notice_api",
      "retrieved_at": "2026-10-11T23:24:32.306399Z"
    }
  ],
  "content": {
    "title": "Scanning electron microscope for quantitative analysis of geological samples",
    "summary": "ETH Zürich is seeking a replacement analytical scanning electron microscope for its Institute of Geochemistry and Petrology. The instrument must support quantitative EDS with reference materials and full matrix correction, include a field emission gun and SE/BSE detectors, and accommodate at least three petrographic thin sections and six one-inch epoxy mounts. Automated quantitative runs longer than 24 hours are required; a cathodoluminescence detector is optional. Delivery has strict lift, floor-loading and crate-size limits. This open, single-lot supply tender accepts electronic bids through Simap. The deadline was extended to 11 November 2026 at 12:00 CET. Detailed specifications and evaluation criteria are in documents that were not accessed.",
    "key_facts": [
      {
        "label": "Purpose",
        "value": "Replace the existing analytical SEM at ETH Zürich’s Institute of Geochemistry and Petrology.",
        "source_ref": "official_ted_api#/description-proc"
      },
      {
        "label": "Core equipment",
        "value": "Field-emission source, SE and high-performance BSE detectors, and large-area EDS spectrometer (≥100 mm²).",
        "source_ref": "official_ted_api#/description-proc"
      },
      {
        "label": "Sample capacity",
        "value": "At least three petrographic thin sections and six one-inch epoxy round mounts.",
        "source_ref": "official_ted_api#/description-proc"
      },
      {
        "label": "Submission deadline",
        "value": "11 November 2026, 12:00 CET (UTC+01:00); extended by two weeks.",
        "source_ref": "official_ted_xml#/ContractNotice/ProcurementProjectLot/TenderingProcess/TenderSubmissionDeadlinePeriod"
      },
      {
        "label": "Delivery limits",
        "value": "Lift-platform vehicle; lifts rated 1,000 kg; floor limit 500 kg/m²; crates at most 90 × 178 × 200 cm.",
        "source_ref": "official_ted_xml#/ContractNotice/ProcurementProject/RealizedLocation/Description"
      }
    ],
    "notice_analysis": {
      "scope": "Supply a replacement analytical scanning electron microscope for geological sample imaging and quantitative EDS analysis at ETH Zürich. The notice specifies detector configuration, sample capacity and long automated runs, plus restrictive delivery conditions.",
      "missing_information": [
        "No estimated contract value, award or contract period is stated in the retrieved notice. Detailed technical specifications and evaluation criteria are in tender documents that were not accessed."
      ],
      "participation_requirements": [
        "The SEM must provide quantitative EDS using reference materials and full matrix correction, and allow automated quantitative runs longer than 24 hours.",
        "The chamber must accommodate at least three petrographic thin sections and six one-inch epoxy mounts.",
        "Delivery requires a lift-platform vehicle; lifts are rated at 1,000 kg, floor loading is limited to 500 kg/m², and crates/deliveries may not exceed 90 × 178 × 200 cm.",
        "Electronic submission via Simap is allowed. Variants are allowed. Detailed qualification and award criteria are referred to tender documents that were not accessed."
      ]
    }
  },
  "facts": {
    "lots": [
      {
        "id": "LOT-0000",
        "title": "Scanning electron microscope for imaging and quantitative analysis in geological samples",
        "source_ref": "official_ted_xml#/ContractNotice/ProcurementProjectLot",
        "description": "Das Institut für Geochemie und Petrologie der ETH Zürich beabsichtigt, sein bestehendes analytisches Rasterelektronenmikroskop (REM) durch ein neues System zu ersetzen. Dieses soll mit einer Feldemissionsquelle, SE- und leistungsstarken BSE-Detektoren sowie einem hochmodernen energiedispersiven Röntgenspektrometer (EDS) mit grossem Detektorbereich (≥ 100 mm²) ausgestattet sein. Wir sind zudem an Angeboten für einen panchromatischen und/oder einen monochromatischen Kathodolumineszenz-Detektor (Einstiegsmodell) interessiert. Hauptziel des Instruments ist die Durchführung präziser und genauer quantitativer EDS-Analysen unter Anwendung eines strengen Quantifizierungsprotokolls, das die Verwendung von Referenzmaterialien sowie eine vollständige Matrixkorrektur (ZAF, phi-rho-z oder gleichwertig) umfasst. Das Gerät muss über eine grosse Probenkammer verfügen, die Platz für mindestens drei petrographische Dünnschliffe und sechs in Epoxidharz eingebettete Rundproben (Durchmesser 1 Zoll) bietet. Zudem soll das System programmierbar sein, um automatisierte quantitative EDS-Analysereihen mit einer Laufzeit von mehr als 24 Stunden zu ermöglichen. Details siehe Pflichtenheft."
      }
    ],
    "schedule": [
      {
        "event": "Notice publication date",
        "value": "1 October 2026",
        "source_ref": "official_ted_xml#/ContractNotice/UBLExtensions/UBLExtension/ExtensionContent/EformsExtension/Publication/PublicationDate"
      },
      {
        "event": "Current tender submission cutoff",
        "value": "11 November 2026 at 12:00 +01:00",
        "source_ref": "official_ted_xml#/ContractNotice/ProcurementProjectLot/TenderingProcess/TenderSubmissionDeadlinePeriod"
      },
      {
        "event": "Deadline amendment",
        "value": "The official notice states the deadline was extended by two weeks.",
        "source_ref": "official_ted_xml#/ContractNotice/UBLExtensions/UBLExtension/ExtensionContent/EformsExtension/Changes/Change/ChangeDescription"
      }
    ],
    "requirements": [
      {
        "text": "Field-emission scanning electron microscope with SE and high-performance BSE detectors and large-area EDS spectrometer of at least 100 mm².",
        "source_ref": "official_ted_api#/description-proc"
      },
      {
        "text": "Quantitative EDS must use reference materials and full matrix correction; automated quantitative runs longer than 24 hours are required.",
        "source_ref": "official_ted_api#/description-proc"
      },
      {
        "text": "The chamber must accommodate at least three petrographic thin sections and six one-inch epoxy round mounts.",
        "source_ref": "official_ted_api#/description-proc"
      },
      {
        "text": "Delivery requires a lift-platform vehicle; lifts are rated to 1,000 kg, floor loading is limited to 500 kg/m², and crates/deliveries may not exceed 90 × 178 × 200 cm.",
        "source_ref": "official_ted_xml#/ContractNotice/ProcurementProject/RealizedLocation/Description"
      },
      {
        "text": "Electronic submission via Simap is allowed and variants are allowed.",
        "source_ref": "official_ted_xml#/ContractNotice/ProcurementProjectLot/TenderingProcess"
      },
      {
        "text": "Qualification and award criteria are referred to tender documentation; those documents were not accessed.",
        "source_ref": "official_ted_xml#/ContractNotice/ProcurementProjectLot/TenderingTerms"
      }
    ],
    "award_criteria": [],
    "estimated_value": null
  },
  "award": null,
  "contractPeriod": null,
  "noticeText": {
    "text": "Titel: Rasterelektronenmikroskop zur Bildgebung und quantitativen Analyse an geologischen Proben\n\nAuftragsbeschreibung:\nDas Institut für Geochemie und Petrologie der ETH Zürich beabsichtigt, sein bestehendes analytisches Rasterelektronenmikroskop (REM) durch ein neues System zu ersetzen. Dieses soll mit einer Feldemissionsquelle, SE- und leistungsstarken BSE-Detektoren sowie einem hochmodernen energiedispersiven Röntgenspektrometer (EDS) mit grossem Detektorbereich (≥ 100 mm²) ausgestattet sein. Wir sind zudem an Angeboten für einen panchromatischen und/oder einen monochromatischen Kathodolumineszenz-Detektor (Einstiegsmodell) interessiert. Hauptziel des Instruments ist die Durchführung präziser und genauer quantitativer EDS-Analysen unter Anwendung eines strengen Quantifizierungsprotokolls, das die Verwendung von Referenzmaterialien sowie eine vollständige Matrixkorrektur (ZAF, phi-rho-z oder gleichwertig) umfasst. Das Gerät muss über eine grosse Probenkammer verfügen, die Platz für mindestens drei petrographische Dünnschliffe und sechs in Epoxidharz eingebettete Rundproben (Durchmesser 1 Zoll) bietet. Zudem soll das System programmierbar sein, um automatisierte quantitative EDS-Analysereihen mit einer Laufzeit von mehr als 24 Stunden zu ermöglichen. Details siehe Pflichtenheft.\n\nAusführungsort und Anlieferung:\nDas Institut für Geochemie und Petrologie der ETH Zürich beabsichtigt, sein bestehendes analytisches Rasterelektronenmikroskop (REM) durch ein neues System zu ersetzen. Dieses soll mit einer Feldemissionsquelle, SE- und leistungsstarken BSE-Detektoren sowie einem hochmodernen energiedispersiven Röntgenspektrometer (EDS) mit grossem Detektorbereich (≥ 100 mm²) ausgestattet sein. Wir sind zudem an Angeboten für einen panchromatischen und/oder einen monochromatischen Kathodolumineszenz-Detektor (Einstiegsmodell) interessiert. Hauptziel des Instruments ist die Durchführung präziser und genauer quantitativer EDS-Analysen unter Anwendung eines strengen Quantifizierungsprotokolls, das die Verwendung von Referenzmaterialien sowie eine vollständige Matrixkorrektur (ZAF, phi-rho-z oder gleichwertig) umfasst. Das Gerät muss über eine grosse Probenkammer verfügen, die Platz für mindestens drei petrographische Dünnschliffe und sechs in Epoxidharz eingebettete Rundproben (Durchmesser 1 Zoll) bietet. Zudem soll das System programmierbar sein, um automatisierte quantitative EDS-Analysereihen mit einer Laufzeit von mehr als 24 Stunden zu ermöglichen. Details siehe Pflichtenheft.\n\nVerfahren: Offenes Verfahren. Auftragsart: Lieferung. Eine Lieferposition (LOT-0000). Elektronische Einreichung über Simap.ch ist zugelassen. Varianten sind zugelassen. Die Eignungskriterien sind in den Ausschreibungsunterlagen festgelegt; die Zuschlagskriterien sind in den Unterlagen definiert.\n\nAngebotsfrist: 11. November 2026, 12:00 Uhr MEZ (UTC+01:00).\nÄnderungsmitteilung: Die Frist für die Angebotsabgabe wurde um zwei (2) Wochen verlängert. Neuer Termin Mittwoch, 11. November 2026 um 12 Uhr mittags via Simap (elektronische Einreichung).",
    "language": "de",
    "source_ref": "official_ted_xml"
  },
  "updates": [
    "The official TED notice states that the submission deadline was extended by two weeks to 11 November 2026 at 12:00 CET, with electronic submission via Simap."
  ],